FDA Emergency Use Authorization
The Solana SARS-CoV-2 Assay is an isothermal Reverse Transcriptase - Helicase-Dependent Amplification (RT-HDA) assay intended for the qualitative detection of nucleic acid from SARS-CoV-2 in nasopharyngeal (NP) and nasal (NS) swab specimens from individuals suspected of COVID-19 by their healthcare provider.
Molecular identification of SARS-CoV-2 occurs by the use of targeting highly conserved regions of the SARS-CoV-2 virus non-structural Polyprotein (pp1ab).
The assay consists of two major steps: (1) specimen preparation, and (2) amplification and detection of target sequences specific to SARS-CoV-2 using isothermal Reverse Transcriptase — Helicase-Dependent Amplification (RT-HDA) in the presence of target-specific fluorescence probes which is performed in the Solana instrument.
Results are displayed on the touchscreen, can be saved to the instrument, printed, and are capable of being sent to the LIS and exported through one of Solana’s five USB ports. Solana SARS-CoV-2 Assay is also supported by the power of Virena®.
Where can I go for updates and more information?
CDC:
General Information - Coronavirus (COVID-19)
Information for Healthcare Professionals
Information for Laboratories
Laboratory Biosafety
Isolation Precautions in Healthcare Settings
Specimen Collection
Infection Control
FDA:
General Information - Coronavirus (COVID-19)
Emergency Use Authorizations
Additional Information - Quidel's EUA COVID-19 Products (Scroll down to the list of EUA products and enter “Quidel” in the search box)
Click Here for Solana SARS-CoV-2 Assay Kit, LYO MMx Cat. #M312
Features & Benefits
Feature | Benefit |
---|---|
RT-HDA technology | Rapid method of isothermal nucleic acid amplification that does not require thermal cycler |
Frozen Master Mix | Easy-to-use format, just thaw Master Mix before use (see Package Insert) |
Small, easy-to-use instrument | Eliminates the need for dedicated molecular space and costly capital equipment, allowing for testing in smaller labs |
CLIA moderately complex | Requires minimal training |
Storage | Process Buffer 2˚C to 8˚C storage, Master Mix -70˚C or below |
Long shelf life | 6 months from date of manufacture |
Room temperature set up | No ice or cooling block required |
Product Specifications
Product Spec Name | Product Spec Data |
---|---|
Sample type | Nasal swabs and nasopharyngeal swabs in transport media* obtained from patients suspected of COVID-19 |
Time to results | Approximately 30 minutes |
Reagent storage conditions | Process Buffer Reaction Tubes 2˚C to 8˚C and Master Mix -70˚C or below |
Controls storage conditions | 2°C to 8°C |
Sample preparation storage conditions | Specimens should be collected, transported, stored, and processed according to CLSI M41-A. Specimens should be stored at 2°C to 8°C until tested. |
Specimens collected in BD/Copan UTM are stable at room temperature (RT), 2°C to 8°C or -70°C or below for up to 4 days. Specimens collected in the CDC Viral Transport Medium should be stored at 2°C to 8°C for up to 72 hours after collection. | |
PPA | Nasal: 97.2% Nasopharyngeal: 100% |
NPA | Nasal: 100% Nasopharyngeal: 96.9% |
*BD/Copan UTM, CDC Viral Transport Media
Ordering Information
Catalog Number | Description | Kit Size / Case Size |
---|---|---|
M313 | Solana SARS-CoV-2 Assay Kit, FRZ MMx | 96 tests, controls |
M956 | Solana SARS-CoV-2 Assay Kit with Nasopharyngeal Swabs and Universal Transport Media | 96 tests, controls, 100 NP swabs, 100 universal transport media |
M958 | Solana SARS-CoV-2 Assay Kit with Nasal Swabs and Universal Transport Media | 96 tests, controls, 100 NS swabs, 100 universal transport media |
M5301 | Solana SARS-CoV-2 Positive Control | Included with kit, additional may be ordered separately |
M5203 | Solana SARS-CoV-2 Negative Control | Included with kit, additional may be ordered separately |